Virtual Instrumentation Design Task of a pressure meter and a spectrum analyser

Authors Avatar

Student: ANDY CRABBE \ IAN MCLOUGHLIN \ PAUL JENNINGS (EPT3)

Lecturer: Dr Deloughry, R

 

THE MANCHESTER METROPOLITAN UNIVERSITY

FACULTY OF SCIENCE AND ENGINEERING

DEPARTMENT OF ENGINEERING AND TECHNOLOGY

Experiment Title:         Virtual Instrumentation Design Task of a pressure meter and a spectrum analyser        

Identity:                I6

Subject Area:                Electronic Instrumentation

Level:                        III

ASSESMENT REPORT

Date 05-12/12/2002                                                     Group Number:EPT3 Group 1

Lecturer: Dr Deloughry, R

Group Members:

1. ANDY CRABBE . (SECTION 3 - Instrument Amplifier)

2. IAN MCLOUGHLIN. (SECTION 4 - Pressure VI design and Accuracy)

3. PAUL JENNINGS. (SECTION 5 - Power Spectrum)

FINALISED AND COMPILED BY IAN MCLOUGHLIN.


Contents

1.        INTRODUCTION        

2.        METHOD        

3. TRANSDUCER AND AMPLIFIER: (Written By ANDY CRABBE)        

3.1.1 Transducer Characteristics        

3.1.2 Effects of IN106A on the Transducer        

3.1.3 Reasons for Effects        

3.2 IN106A Amplifier        

3.2.1 Gain Measurement        

3.2.2 CMRR        

3.3 Overall Instrument Amplifier        

3.3.1 Design of the Buffer Amplifier Values        

3.3.2 Setting up the Buffer Amplifier        

3.3.3 Gain and CMMR measurements        

4 VIRTUAL INSTRUMENT DESIGN: (Written by IAN MCLOUGHLIN)        

4.1 Top Down Design Structure        

4.2 Front Panel and VI Diagram Design        

4.4 Translation of Voltage to Pressure        

4.4 Accuracy        

4.5 Discussion        

4.6 Conclusions        

4.7 References:        

5 SPECTRUM ANALYSER : (Written by PAUL JENNINGS.)        

5.1 Elements of FFT and Spectrum Analyser        

5.2 Results:        

5.3 Conclusions:        


  1. Introduction

The task was to design and build a three op-amp instrumentation amplifier that interfaces between the TSP 410A pressure transducer and a PC computer fitted with an Amplicon PC30 AT data acquisition unit and running National Instruments Labview.  A variable parameter is measured using the pressure transducer and the Labview software is used to measure and convert the output from the transducer and display the result in user friendly manner on the monitor.

The instrumentation amplifier must be such that it gives a maximum output of 5V for a pressure range of 0 to 0.49 Bar.

The instrumentation amplifier itself consists of two stages, namely:

  • Input Buffer Stage (741 op-amps)
  • Differential Amplifier (Texas Instruments INA 106)

The INA 106A differential amplifier consists of precision op-amp and resistors which are laser trimmed for accurate gain and common mode rejection.  Its gain is fixed at G=10 and this was confirmed during the laboratory exercise.

  1. Method

The laboratory work was carried out as described in the handout “Virtual Instrumentation Design Task of a Pressure Meter and a Spectrum Analyser “, and the supplementary worksheet.


3. TRANSDUCER AND AMPLIFIER: (Written By ANDY CRABBE)

3.1.1 Transducer Characteristics

The transducer was a four resistor bridge configuration with a DC supply.  As can be seen from Figure 1, the transducer has a linear characteristic of the form .

Figure 1 - Transducer Characteristic

Where m= Gradient and C= the Intercept.

Thus C=2.968

Therefore,

and hence,

This formula will be required in the Labview virtual instrument to interpret the transducer input.


3.1.2 Effects of IN106A on the Transducer

The INA106 is forms the foundation for many commonly used circuits and provides a precision circuit function without the need for any expensive resistor network.  

With the transducer connected directly to the differential amplifier stage the input voltage from the transducer was measured and recorded.  This measurement was repeated a second time with the differential stage disconnected.  The results are presented in Table 1.

Table 1 - Transducer Output Voltage

It can clearly be seen from the results that the differential amplifier alone causes loading of the transducer.

3.1.3 Reasons for Effects

The pressure transducer is a  wheatstone bridge configuration.  The results given in Table 1 show that connecting the transducer directly to the differential amplifier results in the amplifier loading the transducer.  Consequently, this loading effect will result in errors.  

This loading effect occurs due to the low input impedance of the differential amplifier.  By using a Buffer stage before the differential amplifier, this effect can be overcome.  This is because the Buffer stage will offer high input impedance since it is a non-inverting amplifier at each input.

3.2 IN106A Amplifier

3.2.1 Gain Measurement

Data sheets for the INA 106 differential amplifier state that it has a fixed gain of 10. The differential amplifiers gain is set by the on-chip resistors , where , and so the gain is .  

In order to prove that the gain of the INA 106A amplifier was actually 10, zero volts was applied to terminals A and B (Fig 1), and the offset was adjusted to give zero output.  

Figure 2 - INA 106A Differential Amplifier

A differential input signal was then applied to the amplifier and the gain was measured, the results of which are given in Table 1.

Table 2 - Proof Of INA 106 Gain

3.2.2 CMRR

From the INA106A differential amplifier data  sheet the CMRR is given as 86dB(min) .

To obtain the Common Mode Gain the inputs of the differential stage were initially shorted together and connected to ground.  The output voltage was then checked to ensure it was zero.  These connections were then removed and a voltage was applied to both inputs simultaneously.  With this common 5 volt signal connected, the output voltage was measured at 0.29mV.

Join now!

Thus,

CMRR values above 80dB are considered to be good.

3.3 Overall Instrument Amplifier

3.3.1 Design of the Buffer Amplifier Values

Figure 3 - System Diagram

The overall differential gain of the instrumentation amplifier needs to be

Making

3.3.2 Setting up the Buffer Amplifier

The Buffer stage of the amplifier was then built using the component values calculated above for resistors .  Before connecting the Buffer stage to the Differential stage of the amplifier, the operational amplifiers used in the Buffer stage were adjusted to set any output ...

This is a preview of the whole essay