Forensic Chemistry - Ion Selective Electrode. The objective of experiment is to determine the chloride and nitrate (toxic anions) in samples from the calibration curve

Authors Avatar by desmondseahtm (student)

SINGAPORE POLYTECHNIC

SCHOOL OF CHEMICAL LIFE SCIENCES

Diploma of Applied Chemistry with Pharmaceutical Science

Experiment No. 3

Metal Toxicity

Module Code (CP 4098)

Desmond Seah (P1006812)

Year of Study: Year 2 DACP/FT

Lecturer Jeffrey Whey

AY 11/12


Content Page

Synopsis                                                                                                                    Page

  1. Introduction                                                                                                 1-2  
  2. Theory                                                                                                           2-12  
  3. Procedure                                                                                                     12-13    
  4. Results and calculation                                                                              13-14  
  5. Discussion                                                                                                     14-24  
  6. Conclusion                                                                                                    24  

References

List of Illustration

Synopsis

The objective of experiment is to determine the chloride and nitrate (toxic anions) in samples from the calibration curve. Ion-selective electrodes respond selectivity towards one (or several species) in the presence of others and are used in a wide variety of applications from agricultural, pollution monitoring to dairy products. The potential difference E for an electrode reaction can be given by Nernst equation:

[ E = E0 – RT/nF ln ( 1/ aMn+) ] The ionic activity and concentration are practically identical but in solutions containing many ions, activity and concentration may differ. It is possible to 'fix' the solution so that activity and concentration are equal. This can be done by adding a constant concentration of an inert electrolyte (T.I.S.A.B) to the solutions under test. In order to produce the calibration curve, we have prepared the stock solution for both chloride and nitrate and diluted them accordingly for 3 other concentrations and plotted the potential reading against the concentration in log paper. Hence, we have obtained the concentration of the samples by reading from the calibration curve. Afterwards, we have conducted a t-test and it shows that the unknown correspond to sample D. We have also discussed on some corrective actions that ought to be taken and establishing standard addition method to improve the accuracy of the analysis. All in all, the experiment was a success as all the objectives were met.

1. Objective of Experiment

The objective of experiment is to determine the chloride and nitrate (toxic anions) in water by reading from the calibration curve which involves the calculation and weighing of NaCl and NaNO3 for the preparation of the stock solutions.

 

2. Introduction

Introduction of Ion Selective Electrodes

The term “ion selective electrode” applies to a range of membrane electrodes which respond selectivity towards one (or several species) in the presence of others. The term membrane has associated with the notion of permeability to a species present in test solution.

a) Applications

Ion-selective electrodes are used in a wide variety of applications for determining the concentrations of various ions in aqueous solutions. The following is a list of some of the main areas in which ISEs have been used.

  • Pollution Monitoring: CN, F, S, Cl, NO3 etc., in effluents, and natural waters.
  • Agriculture: NO3, Cl, NH4, K, Ca, I, CN in soils, plant material, fertilisers and feedstuffs.
  • Food Processing: NO3, NO2 in meat preservatives.
  • Salt content of meat, fish, dairy products, fruit juices, brewing solutions.
  • F in drinking water and other drinks.
  • Ca in dairy products and beer
  • K in fruit juices and wine making.
  • Corrosive effect of NO3 in canned foods.
  • Detergent Manufacture: Ca, Ba, F for studying effects on water quality.
  • Paper Manufacture: S and Cl in pulping and recovery-cycle liquors.
  • Explosives: F, Cl, NO3 in explosive materials and combustion products
  • Electroplating: F and Cl in etching baths; S in anodising baths.
  • Biomedical Laboratories: Ca, K, Cl in body fluids (blood, plasma, serum, sweat).
  • F in skeletal and dental studies.

Education and Research: Wide range of applications.

b) Advantages.

  • When compared to many other analytical techniques, Ion-Selective Electrodes are relatively inexpensive and simple to use and have an extremely wide range of applications and wide concentration range.
  • The most recent plastic-bodied all-solid-state or gel-filled models are very robust and durable and ideal for use in either field or laboratory environments.
  • Under the most favourable conditions, when measuring ions in relatively dilute aqueous solutions and where interfering ions are not a problem, they can be used very rapidly and easily (e.g. simply dipping in lakes or rivers, dangling from a bridge or dragging behind a boat).
  • They are particularly useful in applications where only an order of magnitude concentration is required, or it is only necessary to know that a particular ion is below a certain concentration level.
  • They are invaluable for the continuous monitoring of changes in concentration: e.g. in potentiometric titrations or monitoring the uptake of nutrients, or the consumption of reagents.
  • They are particularly useful in biological/medical applications because they measure the activity of the ion directly, rather than the concentration.
  • In applications where interfering ions, pH levels, or high concentrations are a problem, then many manufacturers can supply a library of specialised experimental methods and special reagents to overcome many of these difficulties.
  • With careful use, frequent calibration, and an awareness of the limitations, they can achieve accuracy and precision levels of ± 2 or 3% for some ions and thus compare favourably with analytical techniques which require far more complex and expensive instrumentation.
  • ISEs are one of the few techniques which can measure both positive and negative ions.
  • They are unaffected by sample colour or turbidity.
  •  ISEs can be used in aqueous solutions over a wide temperature range. Crystal membranes can operate in the range 0°C to 80°C and plastic membranes from 0°C to 50°C.

3. Theory

3.1 Potential

When a metal is immersed in a solution containing its own ions, say zinc rod in zinc sulphate solution, a potential difference is established between the metal and the solution.

Mn+ ne-M0

The potential difference E for an electrode reaction can be given by Nernst equation:

E = E0 – RT/nF ln ( 1/ aMn+)

Where R is gas constant, T is absolute temperature,  F is Faraday Constant, n is the valency of ions, aMn+ is the activity of ions in the solution, and E0 is a constant depending upon the metal.

In this equation no term for elemental metal is included in the logarithmic term because it is a pure solid and its activity is one, can be simplified by introducing the known values of R,F and T, and converting natural logarithms to base 10 by multiplying by ~2.3; it then becomes:

E = E0 – 0.0591/n log ( 1/aMn+)

For most purpose in quantitative analysis, it is sufficiently accurate to replace a  aMn+ with [Mn+], the concentration of the metal ion in mol/dm3.

E = E0 – 0.0591/n log( 1/[Mn+] )

This is the simplest form of Nernst equation where E0 is standard electrode potential of the metal, which is constant. Thus, the electrode potential varies linearly with the logarithm the reciprocal of the metal ion concentration. This equation can be used to calculate potential of half-cell if we know the concentration of ions involved in half-cell reaction or calculate the concentration of ions in the solution if we know the potential and standard potential of the half-cell shown above.

3.2 Direct Potentiometry

The direct potentiometry is the method which makes use of the single measured electrode potential to determine the activity of the ion of interest (or concentration of the ion in case of dilute solution). This method is widely used for the determination of pH of the solution (using Hydrogen electrode) or determinations of other ions using different ion-selective electrodes.

In all the electroanalytical methods, two electrodes are required, the potential of one of the electrode be known, constant and is completely independent of the composition of the solution. This electrode is known as Reference electrode.

The second electrode whose potential is dependent on the concentration of the ion to be determined is called indicator electrode. For direct potentiometric measurement, the potential of the cell can be expressed in terms of the potential developed by the indicator electrode and the reference electrode:

ECELL = (EIND - EREF) + EJ

Where EJ is a junction potential.

Consider the indicator electrode whose potential vary with cation activities or concentration, we can write Nernst Equation as derived in 3.1 earlier for half cell reaction.

EIND = EIND0 + (0.0591/n) log aMn+

Substitution of both equations together,

ECELL = (EIND0 + 0.0591/n log aMn+ ) – EREF + EJ

Algebraic manipulation

Log aMn+  =  n(ECELL – K)   /   0.0591

Or pM= -log aMn+ = - n(ECELL – K)   /   0.0591

For anion An- :

pA = n(ECELL – K)   /   0.0591              

Where pM and pA are the negative logarithm of the metal ion activity and an anion respectively. These terms are more general forms of the familiar term pH, K is summation of several constants including standard electrode potential of metal ion or anion, potential reference electrode, junction potential, asymmetrical potential if membrane electrode is involved.

All direct potentiometric methods are based on the last two set of equations. Both equations in terms of ECELL may be written as:

ECELL =  K – 0.0591/n pM

ECELL =  K + 0.0591/n pA

Using last two equations, we can determine concentration of metal ions or cations and anions in terms of pM and pA, respectively. Further, the equation also indicates that for a metal ion-selective electrode, an increase in pM results in decrease in Ecell. But before that constant K should be known to us. As you know K is made up of several constants, including the junction potential, which cannot be measured directly or calculated from theory without assumptions. This problem can be overcome by electrode calibration method using standard solution of the analyte. Though calibration methods are simple, take less time and convenient to the continuous monitoring of pM or pA, but suffer somewhat limited accuracy because of uncertainty in junction potential.

Join now!

In the electrode-calibration method, K is determined by measuring Ecell for one or more standard solution of known pM or pA. The calibration is performed just before the determination of pM or pA for the unknown. There is an assumption that K is unchanged when standard is replaced by the unknown analyte solution. Some time instead of calculating K using emf value of standard solution, a calibration graph is plotted using measured emf of standard solutions against log a (concentration), pM or pA. The sample is then treated in the same way as the standards and the concentration (or ...

This is a preview of the whole essay